XCEL Device Not Detected
This error may appear when the XCEL software is launched or when entering the Capture Screen. Check the following:
- The SphygmoCor XCEL device is physically connected to the computer.
- The device is plugged into a working AC outlet and the power switch is set to ON.
- The Power light is illuminated. (If not, see 'Power Light Is Off' below.)
- The correct communications port is selected in the configuration settings.
- From the Setup screen, select System > Find Module. A successful detection will display 'Module successfully found.'
- If using a USB adapter, confirm that the appropriate USB adapter drivers have been installed on the computer.
If none of the above steps resolve the issue, contact the CONNEQT Health Support Team.